ORIGINAL ARTICLE
Year : 2016  |  Volume : 6  |  Issue : 1  |  Page : 21-25

Correlation of spectral domain optical coherence tomography findings in sub-silicone oil foveal depression space and visual outcome in eyes undergoing silicone oil removal


Department of Retina and Vitreous, Retina Foundation, Gujarat, India

Date of Web Publication13-Feb-2016

Correspondence Address:
Manish Nagpal
Department of Retina and Vitreous , Retina Foundation, Near Shahibaug Underbridge, Rajbhavan Marg, Ahmedabad-380004, Gujarat
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.1016/j.tjo.2015.11.001

Rights and Permissions
  Abstract 


Background/Purpose: To describe small hyper-reflective spherical bodies in sub-silicone oil-foveal depression (SSO-FD) space using spectral domain optical coherence tomography (SD-OCT) and its effect on visual outcomes in eyes undergoing silicone oil removal (SOR).
Methods: This was a prospective interventional comparative study comprising 42 eyes undergoing SOR with clear media. All patients underwent detailed clinical examination and SD-OCT scan of fovea pre-operatively and at 30 days and 90 days postoperatively. Patients were divided into Group A (n = 21) and Group B (n = 21) depending on presence or absence, respectively, of small hyper-reflective spherical bodies in the SSO-FD space in preoperative scans. The findings between SD-OCT and best-corrected visual acuity were correlated and analyzed.
Results: The mean age of patients was 41.9 years (range, 23–60 years) in Group A and 45.6 years (range, 23–60 years) in Group B. Twenty-one eyes showed small hyper-reflective spherical bodies on SD-OCT imaging. These were thought to represent emulsified silicone oil globules trapped in the potential space created by silicone oil meniscus and foveal pit, which is the SSO-FD space. These bodies were absent in all post SOR scans of Group A and Group B. Group A had significant visual improvement (p = 0.0001) after SOR with clearance of these hyper-reflective bodies as compared to Group B(p = 0.356).
Conclusion: We conclude that these small hyper-reflective spherical bodies in the SSO-FD space were most likely emulsified silicone oil globules and correlated with significant visual improvement with their clearance after silicone oil removal.

Keywords: hyper-reflective spherical bodies, silicone oil emulsification, silicone oil removal, spectral domain optical coherence tomography, sub-silicone oil-foveal depression space


How to cite this article:
Nagpal M, Bhatt KJ, Jain P, Taleb EA, Goswami S, Verma A. Correlation of spectral domain optical coherence tomography findings in sub-silicone oil foveal depression space and visual outcome in eyes undergoing silicone oil removal. Taiwan J Ophthalmol 2016;6:21-5

How to cite this URL:
Nagpal M, Bhatt KJ, Jain P, Taleb EA, Goswami S, Verma A. Correlation of spectral domain optical coherence tomography findings in sub-silicone oil foveal depression space and visual outcome in eyes undergoing silicone oil removal. Taiwan J Ophthalmol [serial online] 2016 [cited 2022 Nov 27];6:21-5. Available from: https://www.e-tjo.org/text.asp?2016/6/1/21/204281




  1. Introduction Top


Silicone oil (polydimethylsiloxane) is a linear synthetic polymer composed of repetitive Si–O units and meets all the requirements for intraocular use and can be considered as the ideal material for intraocular tamponade.[1] Cibis et al[2] first reported the use of silicone oil in vitreoretinal surgery in 1962. Later Scott[3] and Zivojnovic et al[4] used this technique in the treatment of complicated retinal detachments with proliferative vitreoretinopathy. In the current era, the use of silicone oil as a surgical tamponade has become a standard technique in the treatment of retinal detachments, especially in proliferative vitreoretinopathy and tractional retinal detachments, severe cases of diabetic retinopathy, endoph-thalmitis, viral retinitis, and ocular trauma.[5]

Emulsification of silicone oil is a well-known phenomenon that is encountered in patients who have had silicone oil tam-ponade for variable periods leading to secondary complications. Although recommendations range from 3 months to 6 months, there is still no definite agreement on the optimal removal time.[6],[7],[8] There is scant scientific knowledge about the in vivo emulsification process of silicone oil intraocularly. Experimental and histopathology studies have shown that silicone oil droplets are deposited within the retinal tissues, optic nerve, and anterior segment structures including the cornea, uveal tissue, and trabecular meshwork.[9],[10],[11],[12]

Chung and Spaide[13] reported the use of first-generation OCT to demonstrate intraretinal silicone oil vacuoles in a single patient who underwent macular hole surgery with internal limiting membrane peeling and temporary silicone oil tamponade. Another study by Errera et al[14] reported the use of SD-OCT in the detection of epiretinal, intraretinal, or subretinal hyper-reflective areas and hypothesized it to be small bubbles of emulsified silicone.

We have noted small spherical bodies in the foveal depression below the silicone oil meniscus showing hyper-reflectivity on SD-OCT scanning. We termed these bodies as hyper-reflective spherical bodies in the subsilicone oil foveal depression (SSO-FD) space. These bodies probably interfere to some extent with central visual acuity. In our prospective comparative study, we compared a group with small hyper-reflective spherical bodies in the SSO-FD space using SD-OCT with a group with no such bodies, and established their correlation with improvement in visual acuity after silicone oil removal.


  2. Methods Top


After Institutional Ethics Committee (Retina foundation, Ahmedabad) approval, 42 eyes of 42 patients who met with all the inclusion and exclusion criteria were enrolled in this study with informed consent. Patients who underwent primary vitrectomy with silicone oil tamponade for rhegmatogenous retinal detachment (macula off), having clear media with settled retina on clinical examination and with a minimum 3 months of post-silicone oil removal follow-up were included in the study. The same medical grade and viscosity silicone oil (1000 centistokes) was used in all patients after vitrectomy. The patients with any complications such as cataract, glaucoma, hyperoleon, or band-shaped keratopathy hampering the OCT scan, or responsible for reduced visual acuity during follow-up of 3 months were excluded. Patients who developed repeat retinal detachments within 3 months after SOR and patients in whom foveal contours were lost due to macular pathology,For each eye, a standard protocol such as epiretinal membrane, scaring, or macular edema, were excluded.

All patients underwent detailed clinical examination including best-corrected visual acuity (BCVA), intraocular pressure measurement and SD-OCT scans before silicone oil removal and at 30 days and 90 days postoperatively. SD-OCT examination was carried out with a Spectralis HRA+OCT device (Heidelberg Engineering, Heidelberg, Germany) equipped with an eye-tracking system for the simultaneous acquisition of near-infrared reflectance (λ = 815 nm), and SD-OCT images with an illumination wavelength of 870 nm and an acquisition speed of 40,000 A scans per second and 7 μ axial resolution. Horizontal and vertical lines as well as volume scans were performed in the morning before silicon oil removal by the surgeon. The patients were followed up at 30 days and 90 days postoperatively.

For each eye, a standard protocol of SD-OCT imaging using the horizontal and vertical scans 30° (8 mm length) and full volume scan containing 19 B scans (30° × 15°) was used. Distance between B scans: 240 µm and 768 A scans was utilized. On follow-up visits, the same areas were scanned using the scanner software by taking progressive reference scan options so as to prevent any observer subjective bias.

Depending on the presence [Figure 1]A, [Figure 2]A, and [Figure 2]C or absence [Figure 3]A of small hyper-reflective spherical bodies in the SSO-FD space in preoperative scans, patients were divided into Group A and Group B. The correlations between preoperative and postoperative SD-OCT findings and BCVA were analyzed. Statistical significance was calculated based on negative ranking using Wilcoxon signed ranks test, Mann–Whitney U test, χ2 and Student t test as appropriate.
Figure 1: (A) Preoperative red-free fundus photograph and spectral domain optical coherence tomography scan in Group A showing: a, foamy glistening sheen; b, hyper-reflective bodies in sub-silicone oil-foveal depression space; with c, intraretinal hyper-reflective bodies; and d, after shadowing/back scattering on retinal layers. (B) Post-silicone oil removal red-free fundus photograph and spectral domain optical coherence tomography scanning in Group A showing: a, absence of foamy sheen; and b, no hyper-reflective bodies in subsilicone oil-foveal depression space. The c, intraretinal hyper-reflective bodies; and d, after shadowing/back scattering on retinal layers are also absent.

Click here to view
Figure 2: (A) Pre-SOR SD-OCT scanning in Group A showing: (a) altered sheen; and (b) hyper-reflective bodies in sub-silicone oil-foveal depression space with back scattering. (B) Post-SOR SD-OCT scan shows absence of altered sheen, hyper-reflective bodies, and back scattering. (C) Pre-SOR SD-OCT scanning in Group A showing hyper-reflective bodies in sub-silicone oil-foveal depression space with back scattering. (D) Post-SOR SD-OCT scan shows absence of hyper-reflective bodies and back scattering. SD-OCT = spectral domain optical coherence tomography; SOR = silicone oil removal.

Click here to view
Figure 3: (A) Pre-SOR and (B) post-SOR spectral domain optical coherence tomography scanning in Group B showing absence of hyper-reflective bodies in sub-silicone oil-foveal depression space. SOR = silicone oil removal.

Click here to view



  3. Results Top


The mean age of patients was 41.9 years (range, 23–60 years) in Group A and 45.6 years (range, 23–60 years) in Group B. The mean duration of silicone oil tamponade was 7.19 ± 1.7 months in Group A and 8.85 ± 3.3 months in Group B. Group A had 66.7% male and 33.3% female patients while Group B had 81% male and 19% female patients [Table 1].
Table 1: Demographic characteristics of study population.

Click here to view


On pre-SOR SD-OCT scans, we found small hyper-reflective spherical bodies in the SSO-FD space in 21 eyes [Figure 1]A, [Figure 2]A, and [Figure 2]C and these cases were kept in Group A, whereas the remaining 21 eyes showed absence of any small hyper-reflective spherical bodies in the SSO-FD space [Figure 3]A and were enrolled in Group B. In addition, 47.6% of patients in Group A showed after-shadowing/back scattering due to accumulated hyper-reflective bodies on the posterior silicone oil meniscus [Figure 1]A and [Figure 2]C on retinal layers. Also, 57.1% of patients in Group A showed intraretinal hyper-reflective bodies [Figure 1]A.

On post-SOR SD-OCT scans, the small hyper-reflective spherical bodies were absent from the SSO-FD space in all cases in Group A [Figure 1]B and [Figure 2]B and Group B [Figure 3]B. The back scattering causing altered reflectivity of the retinal layers [Figure 1]B and [Figure 2]D and intraretinal hyper-reflective bodies were also absent in all the post-SOR scans [Figure 1]B.

The details of mean pre-SOR and post-SOR BCVA at 1 month and 3 months are shown in [Table 2]. The mean improvement in BCVA of Group A at 1 month was 0.173 ± 0.13 and 0.190 ± 0.18 at 3 months, which was highly significant at both 95% and 99% confidence intervals of the difference. There was no improvement seen in mean BCVA in Group B at 1 month, although it improved by 0.114 ± 0.85 at 3 months and was not statistically significant. In Group A, 90.4% of the eyes showed improvement in BCVA while in Group B, only 19% of the eyes showed improvement in BCVA at 3 months post-SOR [Figure 4]. Neither the presence of altered reflectivity (p = 0.751) nor intraretinal hyper-reflective bodies (p = 0.90) correlated significantly with the visual improvement seen in Group A.
Table 2: Mean BCVA (pre-SOR and post-SOR) with mean improvement at 1 month and 3 months.

Click here to view
Figure 4: Mean improvement in best-corrected visual acutiy at 1 month and 3 months post-silicone oil removal.

Click here to view



  4. Discussion Top


Emulsification of silicone oil is a well-known phenomenon that is encountered in patients who have had silicone oil tamponade for variable periods. There is scant scientific knowledge about the in vivo emulsification process of silicone oil intraocularly. However, experimental and histopathological studies have shown that sili-cone oil droplets are deposited within the retinal tissues, optic nerve, and anterior segment structures including the cornea, uveal tissue, and trabecular meshwork.[9],[10],[11],[12] Silicone oil droplets have been seen in the retina, either in enucleated eyes or in eyes with recurrent retinal detachments and having concurrent retinal bi-opsy.[15] Eckardt and associates[15] found single intraretinal macro-phages containing silicone in eyes injected with silicone oil for 2 years and showed that the retina had multiple defects in the inner limiting membrane (ILM). Ohira and coworkers[16] showed that emulsified silicone oil injected into rabbit eyes appeared in the inner retinal layers as early as 1 week after injection. However, advances in OCT technology in recent times have led to increased resolution and better definition of intraocular structures such as the retina, and imaging of these small hyper-reflective spherical bodies and vacuoles is possible.

We have noted small spherical bodies in the foveal depression below the silicone oil meniscus showing hyper-reflectivity on SD-OCT scanning. We hypothesize that these small hyper-reflective spherical bodies, which most likely are emulsified silicone oil globules, are trapped in the potential space created by the silicone oil meniscus and foveal pit, which is the SSO-FD space. To the best of our knowledge, this is the first study to use SD-OCT to describe the presence of hyper-reflective areas in SSO-FD space in eyes with silicone oil tamponade. We also correlated the improvement in BCVA with clearance of these hyper-reflective bodies from the SSO-FD space after SOR, which also has not been reported yet.

A literature review also suggested that these hyper-reflective bodies are possibly emulsified silicone oil globules,[13],[14] or they could be protein aggregates/inflammatory factors/fat.[17],[18] These bodies settles in the fovea and become trapped in the foveal depression. Asaria et al[17] demonstrated that concentrations of basic fibroblast growth factor, interleukin-6 and protein are raised in retro-silicone oil fluid. Jimeno et al[18] demonstrated that cholesterol, fatty acids and derived methyl esters accumulate in intra-vitreal silicone oil used in intraocular tamponade. A recent case report by Welch and de Souza[19] described the possibility of silicone oil microbubble formation and migration within a full thickness macular hole defect contributing to surgical failure. However, there are several observations that suggest that these hyper-reflective spherical bodies are most likely to be emulsified silicone oil.

On review of the literature, we found only two reports of OCT being used as a method to describe residual silicone oil emulsifi-cation after removal. In 2003, Chung and Spaide[13] identified intraretinal small clear silicone oil vacuoles in the macular region where the inner limiting membrane peeling had been performed in a single patient who had undergone macular hole surgery with silicone oil tamponade. They found that these vacuoles were intraretinal cystoid spaces on first-generation OCT. Errera et al[14] reported the use of SD-OCT in the detection of epiretinal, intra-retinal, or subretinal hyper-reflective areas in a series of 11 patients with a history of silicone oil tamponade. They concluded that the hyper-reflective areas are likely to be small bubbles of emulsified silicone. We observed similar hyper-reflective bodies in the SSO-FD space. Errera et al[14] also described identical hyper-reflective spherical bodies in anterior segment SD-OCT after injection of emulsified silicone oil into the model rubber eyes; these bodies were not present in eyes injected with water alone. The difference in reflectivity between our findings those of Chung and Spaide[13] may be attributed to the fact that hyporeflective spaces could represent cystoid macular edema, which can occur after ILM peeling or vitrectomy using silicone oil tamponade. Also, due to technological limitations (1st-generation OCT), previously, multimodal imaging was not performed, and the location of the vacuoles on red-free imaging could not be accurately correlated with the OCT scans. Similar limitations would mean that small, spherical, hyper-reflective bodies would not be resolved.[14]

We found a foamy glistening sheen in red-free photographs [Figure 1]A, [Figure 2]A, and [Figure 2]C, which corresponded to an area of the hyper-reflective spherical SSD-FD space, which cleared along with disappearance of all these hyper-reflective bodies from all eyes after removal of silicone oil [Figure 1]B, [Figure 2]B, and [Figure 2]D, which further supports this hypothesis.

The retinal appearance of retained perfluorocarbon liquid has been described in the literature. Perfluorocarbon liquid bubbles tend to coalesce and are detected as larger hyporeflective entities,[20] distinct from the smaller hyper-reflective entities that we describe. This, to a great extent, confirms that these spherical bodies are not caused by retained perfluorocarbon liquid.

In our clinical scenario, visual acuity of the patients with silicone oil tamponade is reduced due to various reasons such as oil-induced cataract, glaucoma, neuropathy, epiretinal membrane formation, loss of foveal contour, disruption of inner segment/outer segment (IS/OS) junctions, and scarring. Otherwise, the visual acuity remains unchanged in uncomplicated cases. We hypothesize that altered reflectivity of these particles, as noted in SD-OCT scans, created scotoma or some visual impairment. During the process of silicone oil removal, these bodies are washed away from the SSO-FD space and are associated with significant visual improvement. We noted mean BCVA improvement of 0.190 ± 0.18 in Group A and was highly significant at both 95% and 99% confidence intervals of the difference, while Group B did not show any significant visual improvement and remained stable. We also noted visual improvement after SOR in 90.4% of eyes that had hyper-reflective bodies in the SSO-FD space as compared to only 19% of eyes where no such bodies were noted, which further supports our hypothesis.

We found that these hyper-reflective bodies trapped posterior to the silicone oil meniscus, instead of entering the superior part of the eyeball as emulsified silicone oil bubbles, have a natural tendency to enter the superior part.

The limitations of this study were the small sample size and short follow-up. This might have limited the power in detecting other predictors and may have led to insufficiency of the statistical analysis. The 3-month follow-up seems to be appropriate for the evaluation of OCT parameters and visual outcomes, although visual outcomes additionally may improve or decrease, or both, 6 months or even 12 months after surgery. Why these bodies are trapped in the foveal depression, what do they represent, and establishing the correlation between duration of SOR and presence of these hyper-reflective bodies are future areas of research.

In conclusion, our study is believed to be the first to describe the use of SD-OCT imaging to identify in vivo,tiny,hyper-reflective spherical bodies in the SSO-FD space in eyes undergoing SOR. Clearing of these bodies is correlated with visual improvement after SOR. So, the preoperative presence of these spherical bodies on SD-OCT scanning can predict visual improvement and have prognostic significance. SD-OCT represents an important modality in the further investigation of the presence of possible silicone oil emulsification in the SSO-FD space and its clinical significance.

Conflict of Interest: The authors have no conflicts of interest to declare.



 
  References Top

1.
Kreiner CF. Chemical and physical aspects of clinically applied silicones. Dev Ophthalmol. 1987;14:11–19.  Back to cited text no. 1
    
2.
Cibis PA, Becker B, Okun E, Canaan S. The use of liquid silicone in retinal detachment surgery. Arch Ophthalmol. 1962;68:590–599.  Back to cited text no. 2
    
3.
Scott JD. A rationale for the use of liquid silicone. Trans Ophthalmol Soc U K. 1977;97:235–237.  Back to cited text no. 3
    
4.
Zivojnovic R, Mertens DA, Peperkamp E. Liquid silicone in amotio surgery (II). Report on 280 cases – further development of the technic. Klin Monbl Augen- heilkd. 1982;181:444–452.  Back to cited text no. 4
    
5.
Gallemore R, McCuen B. Silicone Oil in Vitreoretinal Surgery. Retina. St Louis: Mosby; 2212–2234.  Back to cited text no. 5
    
6.
Zilis JD, McCuen II BW, de Juan Jr E, et al. Results of silicone oil removal in advanced proliferative vitreoretinopathy. Am J Ophthalmol. 1989;108:15–21.  Back to cited text no. 6
    
7.
Franks WA, Leaver PK. Removal of silicone oil—rewards and penalties. Eye.1991;5:333–337.  Back to cited text no. 7
    
8.
Jiang Y, Li X. The best timing of silicone oil removal. Zhonghua Yan KeZaZhi. 1997;33:39–41.  Back to cited text no. 8
    
9.
Ni C, Wang W, Albert D, et al. Intravitreous silicone injection – histopatho-logical findings in a human eye after 12 years. Arch Ophthalmol. 1983;101:1399–1401.  Back to cited text no. 9
    
10.
Knorr HLJ, Seltsam A, Holbach L, et al. Intraocular silicone oil: a clinicopatho-logical study of 36 enucleated eyes. Ophthalmologie. 1996;93:130–138.  Back to cited text no. 10
    
11.
Parmley VC, Barishak YR, Howes Jr EL, et al. Foreign body giant cell reaction to liquid silicone. Am J Ophthalmol. 1986;101:680–683.  Back to cited text no. 11
    
12.
Heidenkummer HP, Messmer EM, Kampik A. Recurrent vitreoretinal membranes during intravitreal silicone oil tamponade: morphological and immu-nohistochemical investigations. Ophthalmologie. 1996;93:121 –125.  Back to cited text no. 12
    
13.
Chung J, Spaide R. Intraretinal silicone oil vacuoles after macularhole surgery with internal limiting membrane peeling. Am J Ophthalmol. 2003;136: 766–767.  Back to cited text no. 13
    
14.
Errera M-h, Liyanage SE, Elgohary M, et al. Using spectral-domain optical coherence tomography imaging to identify the presence of retinal silicone oil emulsification after silicone oil tamponade. Retina. 2013;33:1567–1573.  Back to cited text no. 14
    
15.
Eckardt C, Nicolai U, Czank M, Schmidt D. Identification of silicone oil in the retina after intravitreal injection. Retina. 1992;12:S17–S22.  Back to cited text no. 15
    
16.
Ohira A, Wilson C, DeJuan E, et al. Experimental retinal tolerance to emulsified silicone oil. Retina. 1991;11:259–265.  Back to cited text no. 16
    
17.
Asaria RHY, Kon CH, Bunce C, et al. Silicone oil concentrates fibrogenic growth factors in the retro-oil fluid. Br J Ophthalmol. 2004;88:1439–1442.  Back to cited text no. 17
    
18.
Jimeno JCP, de la Riia ER, Martinez IF, et al. Lipophilic substances in intraocular silicone oil. Am J Ophthalmol. 2007;143:707–709.  Back to cited text no. 18
    
19.
Welch MJ, De Souza SAM. Silicone oil microbubble found in failed full-thickness macular hole closure. Retinal Cases Brief Rep. 2014;8:132–134.  Back to cited text no. 19
    
20.
Joondeph BC, Nguyen H. Ocular coherence tomography findings with retained submacular perfluoron. Clin Experiment Ophthalmol. 2006;34:85–86.  Back to cited text no. 20
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4]
 
 
    Tables

  [Table 1], [Table 2]


This article has been cited by
1 Synthesis and characteristic analysis of coal dust explosion suppressant based on surface modification of ammonium dihydrogen phosphate with methyl hydrogen-containing silicone oil
Zongqi Liu,Gang Zhou,Shuzheng Song,Qunzhi Meng,Qian Zhang,Wei Liu
Journal of Loss Prevention in the Process Industries. 2020; : 104059
[Pubmed] | [DOI]
2 PREOPERATIVE AND POSTOPERATIVE MACULAR SENSITIVITY ASSESSMENT USING MICROPERIMETRY IN EYES UNDERGOING SILICONE OIL REMOVAL
Manish Nagpal,Jayesh Khandelwal,Ahmed Eltayib,Navneet Mehrotra,Rakesh Juneja
Retina. 2020; 40(8): 1574
[Pubmed] | [DOI]
3 Retinal Layers Measurements following Silicone Oil Tamponade for Retinal Detachment Surgery
Darija Jurišic,Mia Zoric Geber,Ivan Cavar,Dobrila Karlica Utrobicic
Seminars in Ophthalmology. 2017; : 1
[Pubmed] | [DOI]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
1. Introduction
2. Methods
3. Results
4. Discussion
References